TH-DG-150 Temperature & Humidity (Dual) Test Chamber 150L x 2, Temp range: -5 to 100c, 230v 50-60hz.
Environmental test chambers with dual chambers, designed for conducting each different experiment at once.
Stability test in accordance with ICH guidelines Q1A (R2).
Specifications
Model (TH-DG) Dual floor model |
TH-DG-150 |
TH-DG-300 |
TH-DG-400 |
Chamber volume (L / cu ft)
|
150 / 5.30 x 2 chambers
|
288 / 10.17 x 2 chambers
|
384 / 13.56 x 2 chambers
|
Controller
|
PID control (LCD Type)
|
Temperature
|
Range (℃ / ℉)
|
-5 to 100 / 23 to 212 without humidity, 15 to 90 / 59 to 194 with humidity
|
Fluctuation 1) (±℃ / ℉) at 40℃ / 75%RH
|
0.3 / 0.55
|
Variation 1) (±℃ / ℉) at 40℃ / 75%RH
|
0.5 / 0.90
|
Time
|
Heating - without humidity (min.)
|
85
|
Cooling - without humidity (min.)
|
100
|
Humidity
|
Range (%RH)
|
15 to 90
|
Fluctuation 1) (±%RH) at 75%RH / 40℃
|
0.8
|
Variation 1) (±%RH) at 75%RH / 40℃
|
3.0
|
Refrigeration
|
System
|
Air-cooled
|
Refrigerant
|
R-507
|
Water supply
|
Capacity
|
20 / 0.7
|
24 / 0.84 |
Water quality
|
pH6.2 to 7.2 / Electrical conductivity 20㎲ /cm to below
|
No. of wire shelves (standard / max.)
|
2 / 7
|
2 / 8
|
2 / 8
|
Distance of between shelves (mm / inch)
|
60 / 2.4
|
60 / 2.4
|
60 / 2.4
|
Max. load per shelf (kg / lbs)
|
25 / 55.1
|
25 / 55.1
|
25 / 55.1
|
Permitted total load (kg / lbs)
|
75 / 165.3
|
75 / 165.3
|
75 / 165.3
|
Dimension (W×D×H)
|
Interior (mm / inch)
|
600x500x500 / 23.6x19.7x19.7
|
800x600x600 / 31.5x23.6x23.6
|
800x600x800 / 31.5x23.6x31.5
|
Exterior (mm / inch)
|
1055x1320x1725 / 41.5x52.0x67.9
|
1255x1470x1925 / 49.4x57.9x75.8
|
1255x1670x1925 / 49.4x65.7x75.8
|
Viewing windows (mm / inch)
|
200×330 / 7.9×13
|
Net weight (kg / lbs)
|
320 / 705.5
|
385 / 848.8
|
410 / 903.9
|
Cable port
|
One hole provided / Ø50mm (2”)
- Optionally max. two holes available / Ø 50 or 80mm (2 or 3.2”)
|
Interface port
|
Basically RS-232C [RS-485-optional]
|
Electrical requirements (230V, 50-60Hz) |
1ph / 13.0 A x 2ea |
1ph / 16.8 A x 2ea |
Cat. No. |
AAHC6002K |
AAHC6012K |
AAHC6022K |
Warranty (Years) |
2 |
2 |
2 |
1) The set value is set under the conditions of 40°C, 75%RH, technical data according to DIN 12880, IEC 60068. ※ All specifications are under ambient temperature 20℃ (68℉), No load. ※ Permissible environmental conditions are 18 - 30°C (recommended 20°C), 85%RH or less, under 2000m above sea level. ※ The DI water condition is pH 6.2 - 7.2 and the electrical conductivity is less than 20㎲/cm. FDA-registered company.

Performance
- Microprocessor PID control / Temperature calibration / Automatic tuning.
- Advanced performance of low and high temperature and humidity control, achieved by adapting the humidification tank exterior to the chamber, added benefit for long term test.
- Water recycling systems reduce water consumption for longer operations.
- Air-tight door closure with perfect sealed packing.
- Heat-resistant silicone packing completely blocks high heat leakage from the inner chamber.
- Two-point door latch also enhances the tightness of the door closure.
- Energy efficiency is achieved by optimization of the insulation system.
- Even if the unit is operating at max. temperature, the surface of the unit doesn’t allow high temperatures of over 60℃.
Convenience
- Three different volume sizes are available to suit any user's needs.
- Corrosion resistance stainless steel chamber and electric polished humidification tank.
- Water supply during operation is also available for long-term test purposes.
- The direct water system provides users great convenience when performing long-term tests (option).
- Clear observation of samples.
- Tempered glass door with built-in heating wire.
- Energy efficient, long life span LED lighting.
- Wider view, an inner glass door is available for monitoring of samples (option).
- Cable port for introduction of external probes or wires.
- A 50mm diameter cable port comes standard on the left side of the unit.
- Additional 50mm or 80mm diameter cable ports, custom ordered (option).
- Casters for easy mobility during installation or relocation.
- User-friendly LCD controller.
- Colorful interactive touch-type LCD controller.
- The set point temperature and present temperature are clearly shown together on display.
- Sufficient programmable control support.
- Operating progress can be shown in graph mode for better data understanding.
- Automatic start/stop can be programmed for unattended operations.
- The display indicates abnormal temperature changes or power failures.
- Selective bilingual language support.
(Korean / English or English / Chinese, selectable only when placing an order)
- Temperature and other operating parameters can be recorded using a six-channel digital recorder (option).
Computer interface
- RS-485 communication port enables connection up to maximum of 32 units. Simultaneous monitoring and operating of multiple units is an available feature (option).
- Program patterns are simple to set with the use of our software.
- Easy to check the status of real-time operations. Graph data supports easy understanding.
- Storage and documentation of programs.
- Data with saving capabilities in excel and text format.
- Storage of program information, backup of value settings, recovery, and temperature values.
- RS-232 communication port as a standard offer.
Description |
No |
Programmable pattern capacity |
100 |
Pattern repeat time |
999 |
Max. segments per a pattern |
100 |
Available max. segments* |
2000 |
Programmable process time per a segment |
0 to 99hr.
59 min. 59 sec. |
* It is possible to set 100 segments per pattern, but the maximum number of segments is not 10,000 (100 patterns x 100 segments) but 6,000.
Safety
- Built-in door lock with a key.
- Automatic shut off after overheating alarm.
- Low and empty water level alarm.
- Power supply leakage breaker.
- Over current protection.
- Open door alarm.