Lab Companion™ LBV-100 Test Chamber 1000L (Amb +45 to 250c), 380v

Product no.: AAHK1058K

"Our 45th Year of Competitive Pricing"

Ask for a quote for further discounting

Old price $24,750.00

$22,770.00

You save $1,980.00
Price plus sales tax, plus delivery
Delivery weight: 970 lb

A_JeioTech-LC_LOGO

 

 

LBV-100 Heating Test Chamber 1000L, Temp range: (Amb +45 to 250c), Horizontal airflow, 380v 

 

Designed and tested based on 27 temperature measurement points in accordance with international standards.

Horizontal airflow type with intake/damper included as standard.

 

LTV_TEmp_LOGO

 

LTV_Specs

 

LTV_Air_Flow

 

Performance

  • Suitable for a specific use, such as heat treatment and drying process.
  • Microprocessor PID control / Temperature calibration / Automatic tuning.
  • Air-tight door closure with double-sealed packing.
    - Heat-resistant silicone double packing completely blocks high heat leakage from the inner chamber (for LBV).
    - Non-conducting ceramic and heat-resistant silicone packing completely blocks high heat leakage from the inner chamber (for LTV).
    - The two-point door latch also enhances the tightness of the door closure.
  • Energy efficiency is achieved through optimization of the insulation system.
    - If the unit is operating at max. The unit's surface does not allow temperatures above 51℃ (DIN EN 563).
  • Reliable design and production of the door are guaranteed.
    - Evaluates the results after opening the door by 90 degrees and applying a 50kg load at a distance of 30cm from the hinges.

 

Convenience

  • Corrosion-resistant stainless steel inner chamber.
  • Space-efficient and easy-loading chamber design.
    Included two wire shelves for loading large amounts of samples.
    - Heavy load perforated shelving, rated for 40kg also available (option).
  • User-friendly LCD controller.
    - Colorful interactive touch-type LCD controller.
    - Set point temperature and present temperature are clearly shown together on the display.
    - Sufficient programmable control support.
    - Operating progress can be shown in graph mode for better data understanding.
    - Automatic start/stop can be programmed for unattended operations.
    - Display indicates abnormal temperature changes or power failures.
    - Selective bilingual language support.
    (Korean / English or English / Chinese, selectable only when placing an order).
    - Temperature and other operating parameters can be recorded using a six-channel digital recorder (option)
  • Computer interface.
    - RS-485 communication port enables connection up to a maximum of 32 units. Simultaneous monitoring and operating of multiple units is an available feature (option).
    - Program patterns are simple to set with the use of our software.
    - Easy to check the status of real-time operations. Graph data supports easy understanding.
    - Storage and documentation of programs.
    - Data with saving capabilities in Excel and text format.
    - Storage of program information, backup of value settings, recovery, and temperature values.
    - RS-232 communication port as a standard offer.

 

LTV_Segments

Safety

  • User-selectable operation mode after power interruption
    - Automatic run or Manual run.
  • Over-temperature protection.
  • Overcurrent and current-leakage protection (ELCB, earth-leakage circuit breaker).
  • Independent safety fuse.
    - Important parts such as the controller and heaters are highly secured from short circuits. Each heater has an individual fuse to completely break a short circuit.
  • Reversed-phase protection.
    - Electrical connection error indicator helps the user to rapidly fix the connection.
  • Over-current protection of the motor.
    - Automatic shut-off in situations where there is an overcurrent to the motor.
  • Open door warning.
    - The heater and fan stop when the door is in the open state.
  • Electric wiring panel access warning.
    - Automatic stop if the cover of the electric wiring panel is opened.
  • Built-in door lock with a key.

 

 

 

LTV_Temp_Recovery

 

 

 

Product Name Heating Test Chambers (Floor Models)
Model LBV LTV
Airflow Horizontal airflow Horizontal airflow
Chamber volume (L) 125, 253, 420, 720, 1000 125, 253, 420, 720, 1000
Temp. control Operating temp. range (℃/℉) Amb. +45 to 250/ Amb. +81 to 482 Amb. +45 to 350/ Amb. +81 to 662
Fluctuation at 100℃ (±℃/℉) 0.3 / 0.54 0.4, 0.3, 0.5/ 0.72, 0.54, 0.90
Variation at 100℃ (±℃/℉) 0.4, 1.3, 0.5/ 0.72, 2.34, 0.9 0.5, 0.6, 0.7, 0.4/ 0.9, 1.08, 1.26, 0.72
Control Timer Wait on/off Wait on/off
Auto-tuning X X
Program function X X
Computer interface RS-232 port, RS-485 port (optional) RS-232 port, RS-485 port (optional)
Safety Over temp. limit X X
Door open alarm X X
Over-current limit X X
Auto-run (power failure) X X

 

 

LBV_Diagram

 

LBV_Dimensions

 

 

Optional Accessories 

LBV_OP_Accessory

 

 

We also recommend

* Prices plus sales tax, plus delivery

Browse these categories as well: Heating Test Chambers (Floor) LBV & LTV, Catalog